Cardiac implantable electronic devices sense the heart's own electrical activity and deliver therapy when it fails. Permanent pacemakers prevent bradycardia from AV block and sinus node dysfunction; they may be single-chamber, dual-chamber, leadless, or conduction system devices that capture the His bundle or left bundle branch area. Implantable cardioverter-defibrillators (ICDs) detect ventricular tachycardia or fibrillation and treat it with antitachycardia pacing or a shock, preventing sudden death after cardiac arrest and in high-risk cardiomyopathy. Cardiac resynchronisation therapy (CRT) paces both ventricles to correct dyssynchrony in heart failure with LBBB. Every doctor must know the indications, the pacing code, the paced ECG, how a magnet affects each device, MRI and surgical precautions, and the complications: pneumothorax, haematoma, lead displacement, infection, inappropriate shocks and pacing-induced cardiomyopathy. Device infection needs complete system removal, and decisions about deactivating shocks belong in end-of-life planning.
Patient case, mechanism, investigations, management, flashcards and exam-style questions.Open the lesson →

Key facts
- Pacemaker code: chamber paced, chamber sensed, response to sensing (inhibit, trigger, dual), R = rate response; e.g. DDD, VVI, AAI, DDDR
- Main pacing indications: complete and Mobitz II AV block (even without symptoms), symptomatic sinus node dysfunction, alternating bundle branch block, syncope with infra-Hisian disease
- Dual-chamber pacing preserves AV synchrony; VVI(R) or a leadless pacemaker suits permanent AF with slow ventricular response
- Frequent right ventricular pacing (over about 20%) can cause pacing-induced cardiomyopathy; CRT or conduction system pacing avoids it in high-risk patients
- ICD: secondary prevention after VF or unstable VT; primary prevention for LVEF 35% or less in ischaemic cardiomyopathy and selected genetic or non-ischaemic disease
- Subcutaneous ICD avoids intravascular leads but cannot pace for bradycardia or deliver antitachycardia pacing
- Magnet over a pacemaker gives asynchronous fixed-rate pacing; over an ICD it suspends tachyarrhythmia detection and shocks but does not change bradycardia pacing
- Device infection or lead endocarditis: complete extraction of generator and leads plus antibiotics; antibiotics alone fail
Differential diagnosis
| Condition | Clues | Key test |
|---|---|---|
| Single-chamber ventricular pacemaker (VVI/VVIR) | Permanent AF with slow ventricular rate | ECG: ventricular spikes; no atrial lead on X-ray |
| Dual-chamber pacemaker (DDD) | AV block or sinus node dysfunction in sinus rhythm | Atrial and ventricular spikes; two leads on X-ray |
| Leadless pacemaker | High infection risk, poor venous access, dialysis patients | Capsule-shaped device in RV on X-ray |
| Conduction system pacing (His or left bundle branch area) | High expected ventricular pacing, alternative to CRT | Narrow paced QRS; lead deep in septum |
| Transvenous ICD | Needs bradycardia pacing or ATP; secondary or primary prevention | Shock coil on RV lead on X-ray |
| Subcutaneous ICD | Young patients, no pacing need, high infection risk | Lateral generator with parasternal lead, no intracardiac leads |
| CRT-P / CRT-D | HFrEF with LBBB and wide QRS | Three leads including coronary sinus lead |
| Wearable cardioverter-defibrillator | Temporary bridge (e.g. after device extraction) | External vest; no implanted hardware |
Exam pearls
Guideline
2021 ESC guidelines on cardiac pacing and CRT, 2022 ESC ventricular arrhythmia and SCD guideline, 2018 ACC/AHA/HRS bradycardia guideline, 2023 HRS/APHRS/LAHRS physiological pacing guideline, 2025 ESC/EHRA conduction system pacing consensus, 2020 EHRA device infection consensus and 2017 HRS MRI consensus
References
- Glikson M, Nielsen JC, Kronborg MB, et al.. 2021 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J 2021.
- Zeppenfeld K, Tfelt-Hansen J, de Riva M, et al.. 2022 ESC Guidelines for the management of patients with ventricular arrhythmias and the prevention of sudden cardiac death. Eur Heart J 2022.
- Kusumoto FM, Schoenfeld MH, Barrett C, et al.. 2018 ACC/AHA/HRS Guideline on the Evaluation and Management of Patients With Bradycardia and Cardiac Conduction Delay. Circulation 2019.
- Chung MK, Patton KK, Lau CP, et al.. 2023 HRS/APHRS/LAHRS guideline on cardiac physiologic pacing for the avoidance and mitigation of heart failure. Heart Rhythm 2023.
- Moss AJ, Zareba W, Hall WJ, et al.. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med 2002.
- Bardy GH, Lee KL, Mark DB, et al.. Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure. N Engl J Med 2005.
Related topics
By Dr Khalid Seidi (B.Sc., M.D.) · MedEd World Clinical Knowledge library. Original teaching content checked against current guidelines (November 2026); confirm the latest local guidance before clinical use. Not a substitute for clinical judgement.